In-Vivo Evaluation of a Novel Integrated Pediatric Pump Lung in a 30-Day Ovine Animal Model.

阅读:5
作者:Leibowitz Joshua L, Awad Morcos A, Han Dong, Shah Aakash, Sun Wenji, Zhang Jiafeng, Griffith Bartley P, Wu Zhongjun J
Recently there has been increased use of mechanical circulatory support in pediatric patients as a bridge to cardiopulmonary recovery or transplantation. However, there are few devices that are optimized and approved for use in pediatric patients. We designed and prototyped a novel integrated pediatric pump lung (PPL) that underwent 30 day in-vivo testing in seven juvenile Dorset Hybrid sheep. Devices were implanted in a right atrial to pulmonary artery configuration. Six of seven sheep survived with a device functioning for 25-35 days. The device flow rate was maintained at 2.08 ± 0.34 to 2.54 ± 0.16 L/min with oxygen transfer of 109.8 ± 24.8 to 151.2 ± 26.2 ml/min over the study duration. Aside from a postoperative drop in hematocrit, all hematologic and blood chemistry test values returned to normal ranges after 1-2 weeks postoperatively. Similarly, lactate dehydrogenase increased postoperatively and returned to baseline. In two sheep, there were early device failures due to oxygenator thrombosis on postoperative days zero and five; they then had oxygenator exchanges with subsequent devices performing stably for 30 days. This study demonstrated that the integrated PPL device exhibited stable performance and acceptable biocompatibility in a 30 day ovine model.