Optimum Diffraction-Corrected Frequency-Shift Estimator of the Ultrasonic Attenuation Coefficient.

超声衰减系数的最佳衍射校正频移估计器

阅读:8
作者:Samimi Kayvan, Varghese Tomy
The ultrasonic attenuation coefficient is an important parameter that has been studied extensively in Quantitative Ultrasound and Tissue Characterization. There are various methods described in the literature that estimate this parameter by measuring either spectral difference (i.e., decay) or spectral shift of the backscattered echo signal. Under ideal conditions, i.e., in the absence of abrupt changes in tissue backscattering, Spectral Difference methods can produce estimates with high accuracy and precision. On the other hand, diffraction-corrected Spectral Shift methods (e.g., the Hybrid method) are better suited for application in practical settings using clinical ultrasound scanners. However, current Spectral Shift methods use inefficient frequency shift estimators that ultimately degrade the quality of attenuation coefficient estimates. In this paper, a probabilistic model of the backscattered radiofrequency (RF) echo is used to derive the Cramér-Rao lower bound (CRLB) on estimation variance of the spectral centroid. Next, an efficient correlation-based shift estimator is presented that achieves the CRLB. Used in conjunction with a well-characterized reference phantom to correct for diffraction and other system-related effects, this estimator greatly improves the accuracy and precision of Spectral- Shift attenuation estimation. A theoretical analysis of this method is provided, and its performance is quantitatively compared with that of the Hybrid method using simulated and experimental phantom studies. A minimum of 3-fold reduction in the standard deviation of attenuation coefficient estimates is observed using the new method.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。