This study provides a detailed comparison of three optical sensing approaches for colourimetric bio/chemical detection, focusing on cost, scalability, and performance. We examine laboratory-grade spectrophotometry, portable camera-based imaging, and low-cost LED photometry using Paired Emitter-Detector Diode (PEDD) charge-discharge methodology. Our findings reveal that while the LED-based PEDD system outperforms the other two methods in key sensory metrics-such as sensitivity, resolution, and limit of detection-its cost-effectiveness and scalability make it a promising solution for widespread industrial and field applications. Compared to the spectrophotometer, the LED/PEDD approach demonstrates improvements in measurement range (Ã16.39), dynamic range (Ã147.06), accuracy (Ã1.79), and sensitivity (Ã107.53). The results highlight the potential for industrial-scale adoption of LED photometry, especially for cost-effective applications in bio/chemical sensing sectors.
A Comparative Study of Optical Sensing Methods for Colourimetric Bio/Chemical Detection: Cost, Scale, and Performance.
比色生物/化学检测的光学传感方法比较研究:成本、规模和性能
阅读:4
作者:Fay Cormac D, Wu Liang, Perez de Vargas Sansalvador Isabel M
| 期刊: | Sensors | 影响因子: | 3.500 |
| 时间: | 2025 | 起止号: | 2025 Jun 20; 25(13):3850 |
| doi: | 10.3390/s25133850 | ||