CHAMP: A Centrifugal Microfluidics-Based CRISPR/Cas12b-Combined Real-Time LAMP One-Pot Method for Mycoplasma pneumoniae Infection Diagnosis.

阅读:9
作者:Jing Wenwen, Zhang Tong, Min Xiangyang, Li Xin, Jin Kai, Feng Meng, Sui Guodong, Luo Liulin, Cheng Xunjia
The Mycoplasma pneumoniae outbreak poses health risks to community residents. However, it still has limitations for current clinical diagnostic methods (qPCR nucleic acid assay or IgM immunoassay), including specialized handling, expensive equipment, prolonged turnaround time, and false positives and negatives, highlighting the need to improve clinical diagnostic methods. Herein, we present a novel centrifugal microfluidics-based method for rapidly diagnosing M. pneumoniae infections (CHAMP system). This user-friendly method combines CRISPR/Cas12b and real-time loop-mediated isothermal amplification (LAMP) in a one-pot reaction, offering high sensitivity, specificity, and simplicity for methodology. By adding fully automated nucleic acid magnetic bead-extracted samples to a prepackaged centrifugal microfluidics chip, 48 samples can be automated tested simultaneously within 15 to 60 min at 60 °C. 427 clinical nasopharyngeal swab specimens were used for validation, demonstrating good positive and negative predictive values and good diagnostic sensitivity, specificity, and significant time savings. This method is particularly suitable for detecting low nucleic acid copies of M. pneumoniae samples.