A hemagglutinin stabilized stem nanoparticle (HA-SS-np) that is designed to provide broad protection against influenza is being developed as a potential vaccine. During an early formulation screening study, reducing gel (rCGE) analysis indicated product degradation in a few candidate buffers at the first-week accelerated stability point, whereas no change was shown in the size exclusion chromatography (SEC) measurement. A LC-MS workflow was therefore applied to investigate the integrity of this large HA-SS-np vaccine molecule (ââ1Â MDa). Application of LC-MS was critical to rationalize the conflicting results from the rCGE and SEC assays and led to the discovery that (1) an unexpected sequence clipping in the HA-SS-np subunits occurred, explaining the atypical reducing gel profile, and (2) an undisrupted disulfide bond held the two fragments together, explaining the unchanged SEC profile. This analytical case study led to a formulation buffer redesign, which mitigated the issue.
Using LC-MS to Identify Clipping in Self-Assembled Nanoparticles During Vaccine Development.
阅读:21
作者:Schneck Nicole A, Ivleva Vera B, Rosales-Zavala Erwin, Wang Xiangchun, Gollapudi Deepika, Cooper Jonathan W, Lei Q Paula
| 期刊: | Journal of the American Society for Mass Spectrometry | 影响因子: | 2.700 |
| 时间: | 2019 | 起止号: | 2019 Dec;30(12):2576-2579 |
| doi: | 10.1007/s13361-019-02318-z | ||