Validation and measurement uncertainty evaluation of the GC-MS method for the direct determination of hexamethylenetetramine from foods

对直接测定食品中六亚甲基四胺的GC-MS方法进行验证和测量不确定度评价

阅读:2

Abstract

A simple GC-MS method has been developed and validated for the direct determination of hexamethylenetetramine (HMT). The separation of HMT was performed using a MXT-1 column. The calibration curve was linear over the concentration range 0.1-25 μg/mL, with a good correlation coefficient (r =0.9996). The recoveries of HMT from foods spiked at 1, 5, and 10 μg/g ranged from 91.7% to 115.2%. Intra-day (n=5) and inter-day (n=5) precision were less than 7%. The limit of detection and the limit of quantification of the method were 0.05 and 0.15 μg/mL, respectively. The uncertainties associated with food matrix and calibration contributed most to the overall expanded uncertainty. The method validation data indicated that quantitative method could be applied to the direct determination of non-hydrolyzed HMT in foods.