Performance verification of the new fully automated Aquios flow cytometer PanLeucogate (PLG) platform for CD4-T-lymphocyte enumeration in South Africa

在南非对新型全自动Aquios流式细胞仪PanLeucogate (PLG)平台进行CD4-T淋巴细胞计数的性能验证

阅读:1

Abstract

BACKGROUND: The National Health Laboratory Service (NHLS) offers wide-scale CD4 testing through a network of laboratories in South Africa. A new "load and go" cytometer (Aquios CL, Beckman Coulter), developed with a PLG protocol, was validated against the predicate PLG method on the Beckman Coulter FC500 MPL/CellMek platform. METHODS: Remnant routine EDTA blood CD4 reference results were compared to results from two Aquios/PLG instruments (n = 205) and a further n = 1885 samples tested to assess daily testing capacity. Reproducibility was assessed using ImmunotrolTM and patient samples with low, medium, high CD4 counts. Data was analyzed using GraphPad software for general statistics and Bland-Altman (BA) analyses. The percentage similarity (%Sim) was used to measure the level of agreement (accuracy) of the new platform versus the predicate and variance (%SimCV) reported to indicate precision of difference to predicate. RESULTS: 205 samples were tested with a CD4 count range of 2-1228 cells/μl (median 365cells/μl). BA analysis revealed an overall -40.5±44.0cells/μl bias (LOA of 126.8 to 45.8cells/μl) and %Sim showing good agreement and tight precision to predicate results (94.83±5.39% with %SimCV = 5.69%). Workflow analysis (n = 1885) showed similar outcomes 94.9±8.9% (CV of 9.4%) and 120 samples/day capacity. Excellent intra-instrument reproducibility was noted (%Sim 98.7±2.8% and %SimCV of 2.8%). 5-day reproducibility using internal quality control material (Immunotrol™) showed tight precision (reported %CV of 4.69 and 7.62 for Normal and Low material respectively) and instrument stability. CONCLUSION: The Aquios/PLG CD4 testing platform showed clinically acceptable result reporting to existing predicate results, with good system stability and reproducibility with a slight negative but precise bias. This system can replace the faded XL cytometers in low- to medium volume CD4 testing laboratories, using the standardized testing protocol, with better staff utilization especially where technical skills are lacking. Central monitoring of on-board quality assessment data facilitates proactive maintenance and networked instrument performance monitoring.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。