Sources of Uncertainty in a DVM-Based Measurement System for a Quantized Hall Resistance Standard

基于数字电压表的量化霍尔电阻标准测量系统中的不确定性来源

阅读:1

Abstract

Transportable 10 kΩ standard resistors have become fairly widespread in industrial, university, and government standards laboratories because of their low temperature coefficient of resistance, case of transportation, and convenient value. The values of these resistors, however, tend to drift with time, requiring periodic recalibration against an invariant standard such as the quantized Hall resistance. The availability of a simple, inexpensive measurement system for calibrating 10 kΩ resistors against such an invariant standard would be of great benefit to primary standards laboratories. This paper describes a simple automated measurement system that uses a single, high accuracy, commercially available digital voltmeter (DVM) to compare the voltages developed across a 10 kΩ standard resistor and a quantized Hall resistor when the same current is passed through the two devices. From these measurements, the value of the 10 kΩ standard resistor is determined. The sources of uncertainty in this system are analyzed in detail and it is shown that it is possible to perform calibrations with relative combined standard uncertainties less than 1×10(-7) (0.1 ppm).