Automated detection of shadow artifacts in optical coherence tomography angiography

光学相干断层扫描血管造影术中阴影伪影的自动检测

阅读:3

Abstract

Frequently, when imaging retinal vasculature with optical coherence tomography angiography (OCTA) in diseased eyes, there are unavoidable obstacles to the propagation of light such as vitreous floaters or the pupil boundary. These obstacles can block the optical coherence tomography (OCT) beam and impede the visualization of the underlying retinal microcirculation. Detecting these shadow artifacts is especially important in the quantification of metrics that assess retinal disease progression because they might masquerade as regional perfusion loss. In this work, we present an algorithm to identify shadowed areas in OCTA of healthy subjects as well as patients with diabetic retinopathy, uveitis and age-related macular degeneration. The aim is to exclude these areas from analysis so that the overall OCTA parameters are minimally affected by shadow artifacts.