Measurement of skin dose from cone-beam computed tomography imaging

利用锥形束计算机断层扫描成像测量皮肤剂量

阅读:1

Abstract

OBJECTIVE: To measure surface skin dose from various cone-beam computed tomography (CBCT) scanners using point-dosimeters. MATERIALS & METHODS: A head anthropomorphic phantom was used with nanoDOT optically stimulated luminescence (OSL) dosimeters (Landauer Corp., Glenwood, IL) attached to various anatomic landmarks. The phantom was scanned using multiple exposure protocols for craniofacial evaluations in three different CBCT units and a conventional x-ray imaging system. The dosimeters were calibrated for each of the scan protocols on the different imaging systems. Peak skin dose and surface doses at the eye lens, thyroid, submandibular and parotid gland levels were measured. RESULTS: The measured skin doses ranged from 0.09 to 4.62 mGy depending on dosimeter positions and imaging systems. The average surface doses to the lens locations were ~4.0 mGy, well below the threshold for cataractogenesis (500 mGy). The results changed accordingly with x-ray tube output (mAs and kV) and also were sensitive to scan field of view (SFOV). As compared to the conventional panoramic and cephalometric imaging system, doses from all three CBCT systems were at least an order of magnitude higher. CONCLUSIONS: Peak skin dose and surface doses at the eye lens, thyroid, and salivary gland levels measured from the CBCT imaging systems were lower than the thresholds to induce deterministic effects. However, our findings do not justify the routine use of CBCT imaging in orthodontics considering the lifetime-attributable risk to the individual.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。