High variability in physician estimations of flow-diverting stent deployment versus PreSize Neurovascular software simulation: a comparison study

医生对血流导向支架植入的估计与 PreSize Neurovascular 软件模拟结果存在显著差异:一项对比研究

阅读:1

Abstract

BACKGROUND: Physician variablity in preoperative planning of endovascular implant deployment and associated inaccuracies have not been documented. This study aimed to quantify the variability in accuracy of physician flow diverter (FD) planning and directly compares it with PreSize Neurovascular (Oxford Heartbeat Ltd) software simulations. METHODS: Eight experienced neurointerventionalists (NIs), blinded to procedural details, were provided with preoperative 3D rotational angiography (3D-RA) volumetric data along with images annotated with the distal landing location of a deployed Surpass Evolve (Stryker Neurovascular) FD from 51 patient cases. NIs were asked to perform a planning routine reflecting their normal practice and estimate the stent's proximal landing using volumetric data and the labeled dimensions of the FD used. Equivalent deployed length estimation was performed using PreSize software. NI- and software-estimated lengths were compared with postprocedural observed deployed stent length (control) using Bland-Altman plots. NI assessment agreement was assessed with the intraclass correlation coefficient (ICC). RESULTS: The mean accuracy of NI-estimated deployed FD length was 81% (±15%) versus PreSize's accuracy of 95% (±4%), demonstrating significantly higher accuracy for the software (p<0.001). The mean absolute error between estimated and control lengths was 4 mm (±3.5 mm, range 0.03-30.2 mm) for NIs and 1 mm (±0.9 mm, range 0.01-3.9 mm) for PreSize. No discernable trends in accuracy among NIs or across vasculature and aneurysm morphology (size, vessel diameter, tortuousity) were found. CONCLUSIONS: The study quantified experienced physicians' significant variablity in predicting an FD deployment with current planning approaches. In comparison, PreSize-simulated FD deployment was consistently more accurate and reliable, demonstrating its potential to improve standard of practice.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。