Liver Cancer Detection by a Simple, Inexpensive and Effective Immunosensor with Zinc Oxide Nanoparticles

利用氧化锌纳米粒子构建简单、廉价、高效的免疫传感器检测肝癌

阅读:1

Abstract

Regular monitoring of blood α-fetoprotein (AFP) and/or carcino-embryonic antigen (CEA) levels is important for the routine screening of liver cancer. However, AFP and CEA have a much lower specificity than des-γ-carboxyprothrombin (DCP) to detect liver cancer. Therefore, the study reported here was designed, to develop a screen-printed DCP immunosensor incorporating zinc oxide nanoparticles, for accurate determination of DCP. The designed immunosensor shows low detection limits for the detection of DCP: 0.440 ng/mL (based on impedance measurement), 0.081 ng/mL (based on real part of impedance measurement) and 0.078 ng/mL (based on imaginary part of impedance measurement), within the range of 3.125 ng/mL to 2000 ng/mL. In addition, there was little interference to DCP determination by molecules such as Na⁺, K⁺, Ca(2+), Cl(-), glucose, urea, and uric acid. It is therefore concluded that the DCP immunosensor developed and reported here is simple, inexpensive and effective, and shows promise in the rapid screening of early-stage liver cancer at home with a point-of-care approach.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。