Reliability and limits of transport-ventilators to safely ventilate severe patients in special surge situations

在特殊紧急情况下,转运呼吸机在安全通气重症患者方面的可靠性和局限性

阅读:1

Abstract

BACKGROUND: Intensive Care Units (ICU) have sometimes been overwhelmed by the surge of COVID-19 patients. Extending ICU capacity can be limited by the lack of air and oxygen pressure sources available. Transport ventilators requiring only one O(2) source may be used in such places. OBJECTIVE: To evaluate the performances of four transport ventilators and an ICU ventilator in simulated severe respiratory conditions. MATERIALS AND METHODS: Two pneumatic transport ventilators, (Oxylog 3000, Draeger; Osiris 3, Air Liquide Medical Systems), two turbine transport ventilators (Elisee 350, ResMed; Monnal T60, Air Liquide Medical Systems) and an ICU ventilator (Engström Carestation-GE Healthcare) were evaluated on a Michigan test lung. We tested each ventilator with different set volumes (Vt(set) = 350, 450, 550 ml) and compliances (20 or 50 ml/cmH(2)O) and a resistance of 15 cmH(2)O/l/s based on values described in COVID-19 Acute Respiratory Distress Syndrome. Volume error (percentage of Vt(set)) with P(0.1) of 4 cmH(2)O and trigger delay during assist-control ventilation simulating spontaneous breathing activity with P(0.1) of 4 cmH(2)O and 8 cmH(2)O were measured. RESULTS: Grouping all conditions, the volume error was 2.9 ± 2.2% for Engström Carestation; 3.6 ± 3.9% for Osiris 3; 2.5 ± 2.1% for Oxylog 3000; 5.4 ± 2.7% for Monnal T60 and 8.8 ± 4.8% for Elisee 350. Grouping all conditions (P(0.1) of 4 cmH(2)O and 8 cmH(2)O), trigger delay was 50 ± 11 ms, 71 ± 8 ms, 132 ± 22 ms, 60 ± 12 and 67 ± 6 ms for Engström Carestation, Osiris 3, Oxylog 3000, Monnal T60 and Elisee 350, respectively. CONCLUSIONS: In surge situations such as COVID-19 pandemic, transport ventilators may be used to accurately control delivered volumes in locations, where only oxygen pressure supply is available. Performances regarding triggering function are acceptable for three out of the four transport ventilators tested.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。