Reliability of respiratory pressure measurements in ventilated and non-ventilated patients in ICU: an observational study

重症监护室机械通气和非机械通气患者呼吸压力测量可靠性的观察性研究

阅读:1

Abstract

BACKGROUND: Assessment of maximum respiratory pressures is a common practice in intensive care because it can predict the success of weaning from ventilation. However, the reliability of measurements through an intubation catheter has not been compared with standard measurements. The aim of this study was to compare maximum respiratory pressures measured through an intubation catheter with the same measurements using a standard mouthpiece in extubated patients. METHODS: A prospective observational study was carried out in adults who had been under ventilation for at least 24 h and for whom extubation was planned. Maximal respiratory pressure measurements were carried out before and 24 h following extubation. RESULTS: Ninety patients were included in the analyses (median age: 61.5 years, median SAPS2 score: 42.5 and median duration of ventilation: 7 days). Maximum respiratory pressures measured through the intubation catheter were as reliable as measurements through a standard mouthpiece (difference in maximal inspiratory pressure: mean bias = - 2.43 ± 14.43 cmH(2)O and difference in maximal expiratory pressure: mean bias = 1.54 ± 23.2 cmH(2)O). CONCLUSION: Maximum respiratory pressures measured through an intubation catheter were reliable and similar to standard measures. Clinical trial registration Retrospectively Registered in ClinicalTrials.gov (NCT02363231).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。