Performance characteristics of the Cavidi ExaVir viral load assay and the ultra-sensitive P24 assay relative to the Roche Monitor HIV-1 RNA assay

Cavidi ExaVir病毒载量检测和超灵敏P24检测相对于Roche Monitor HIV-1 RNA检测的性能特征

阅读:1

Abstract

BACKGROUND: The Cavidi viral load assay and the ultra-sensitive p24 antigen assay (Up24 Ag) have been suggested as more feasible alternatives to PCR-based HIV viral load assays for use in monitoring patients infected with HIV-1 in resource-limited settings. OBJECTIVES: To describe the performance of the Cavidi ExaVir Load™ assay (version 2.0) and two versions of the Up24 antigen assay and to characterize their agreement with the Roche Monitor HIV-1 RNA assay (version 1.5). STUDY DESIGN: Observational study using a convenience sample of 342 plasma specimens from 108 patients enrolled in two ACTG clinical trials to evaluate the performance characteristics of the Up24 Ag assay using two different lysis buffers and the Cavidi ExaVir Load™ assay. RESULTS: In analysis of agreement with the Roche assay, the Cavidi assay demonstrated superiority to the Up24 Ag assays in accuracy and precision, as well as sensitivity, specificity, and positive and negative predictive values for HIV-1 RNA ≥ 400, ≥ 1000 and ≥ 5000 copies/mL. Logistic performance curves indicated that the Cavidi assay was superior to the Up24 assays for viral loads greater than 650 copies/mL. CONCLUSIONS: The results suggest that the Cavidi ExaVir Load assay could be used for monitoring HIV-1 viral load in resource-limited settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。