Soft, adhesive and conductive composite for electroencephalogram signal quality improvement

用于改善脑电图信号质量的柔软、粘合和导电复合材料

阅读:15
作者:Jeong E Jin, Seohyeon Kim, Hyeji Yu, Keyong Nam Lee, Young Rag Do, Seung Min Lee

Abstract

Since electroencephalogram (EEG) is a very small electrical signal from the brain, it is very vulnerable to external noise or motion artifact, making it difficult to measure. Therefore, despite the excellent convenience of dry electrodes, wet electrodes have been used. To solve this problem, self-adhesive and conductive composites using carbon nanotubes (CNTs) in adhesive polydimethylsiloxane (aPDMS), which can have the advantages of both dry and wet electrodes, have been developed by mixing them uniformly with methyl group-terminated PDMS. The CNT/aPDMS composite has a low Young's modulus, penetrates the skin well, has a high contact area, and excellent adhesion and conductivity, so the signal quality is enhanced. As a result of the EEG measurement test, although it was a dry electrode, results comparable to those of a wet electrode were obtained in terms of impedance and motion noise. It also shows excellent biocompatibility in a human fibroblast cell test and a week-long skin reaction test, so it can measure EEG with high signal quality for a long period of time.