Decay-Associated Fluorescence for Boron Determination in Uranium-Based Nuclear Fuels

利用衰变相关荧光测定铀基核燃料中的硼

阅读:7
作者:Poonam Verma

Abstract

Herein is reported a novel as well as simple, sensitive, and cost-effective method for determination of boron by time-resolved fluorescence spectrometry in uranium-based nuclear fuels. Boron is complexed with fluorescent ligand chromotropic acid, and the complex formed is a measure of boron. Since the steady-state fluorescence spectra of excess ligand and complex are overlapping, the developed method emphasizes the power of time resolution. The signatory fluorescence decay times of ligand and complex are employed to derive their decay-associated spectra (DAS) and, thereby, spectroscopically eliminate the high background of ligand fluorescence. The calibration plot has a wide linear dynamic range of 5-100 ppb with r2 better than 0.998. Precision is better than 5% at the 10 ppb level and 4% at the 50 ppb level (n = 9). The detection limit is 1.5 ppb, and recovery of spiked boron (25 ppb) from the uranium samples was better than 94%. The developed method was validated by analyzing U3O8-based ILCE Standards and applied to enriched uranium fuel samples. The main advantage of the developed method is a reduction in sample size requirement due to better sensitivity and selectivity. This in turn reduces the load of uranium recovery from analytical waste, especially in the case of enriched uranium samples. Additionally, it eliminates the need of organic solvents/medium.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。