Commercial and Custom Quartz Tuning Forks for Quartz Enhanced Photoacoustic Spectroscopy: Stability under Humidity Variation

用于石英增强光声光谱的商用和定制石英音叉:湿度变化下的稳定性

阅读:1

Abstract

This work investigates the behavior of commercial and custom Quartz tuning forkss (QTF) under humidity variations. The QTFs were placed inside a humidity chamber and the parameters were studied with a setup to record the resonance frequency and quality factor by resonance tracking. The variations of these parameters that led to a 1% theoretical error on the Quartz Enhanced Photoacoustic Spectroscopy (QEPAS) signal were defined. At a controlled level of humidity, the commercial and custom QTFs present similar results. Therefore, commercial QTFs appear to be a very good candidates for QEPAS as they are also affordable and small. When the humidity increases from 30 to 90 %RH, the variations in the custom QTFs' parameters remain suitable, while commercial QTFs show unpredictable behavior.