Assessment of the clinical efficacy of the heart spectrum blood pressure monitor for diagnosis of atrial fibrillation: An unblinded clinical trial

评估心谱血压监测仪在诊断房颤中的临床疗效:一项非盲临床试验

阅读:3

Abstract

Atrial fibrillation (AF) is the most common arrhythmia. The most common diagnostic method, 12-lead electrocardiogram (ECG), can record episodes of arrhythmia from which the type and severity can be determined. The Heart Spectrum Blood Pressure Monitor (P2; OSTAR Meditech Corp., New Taipei City, Taiwan) is used to measure cardiovascular pressure change with fast Fourier transform (FFT) analysis to obtain heart rate frequency variability and accurate blood pressure data. We compared the diagnostic efficacy of the Heart Spectrum Blood Pressure Monitor to a 12-lead ECG (gold standard) for patients with AF. Three measurement methods were used in this study to analyze the heart index and compare the results with simultaneous 12-lead ECG: blood pressure; mean arterial pressure, which was calculated from individual blood pressure as a constant pressure; and a constant pressure of 60 mmHg. The physician used a 12-lead ECG and the Heart Spectrum Blood Pressure Monitor simultaneously. The Heart Spectrum Blood Pressure Monitor used FFT analysis to diagnose AF, and the findings were compared to the 12-lead ECG readings. This unblinded clinical trial was conducted in the emergency department of Taipei Medical University Hospital. Twenty-nine subjects with AF and 33 without AF aged 25 to 97 y (mean, 63.5 y) were included. Subjects who were exposed to high-frequency surgical equipment during testing, those with cardiac pacemakers or implantable defibrillators, and pregnant women were excluded. The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were 97%, 97%, 97%, and 97%, respectively, for method 1; 90%, 100%, 100%, and 91%, respectively, for method 2; and 100%, 94%, 94%, and 100%, respectively, for method 3. The sensitivity, specificity, PPV, and NPV for both methods ranged between 90% and 100%, indicating that the Heart Spectrum Blood Pressure Monitor can be effectively applied for AF detection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。