Trace Voltammetric Determination of Lead at a Recycled Battery Carbon Rod Electrode

利用痕量伏安法测定回收电池碳棒电极上的铅含量

阅读:1

Abstract

Carbon rod electrodes (CREs) were obtained from recycled zinc⁻carbon batteries and were used without further modification for the measurement of trace concentrations of lead (Pb). The electrochemical behavior of Pb at these electrodes in a variety of supporting electrolytes was investigated by cyclic voltammetry. The anodic peaks obtained on the reverse scans were indicative of Pb being deposited as a thin layer on the electrode surface. The greatest signal⁻to⁻noise ratios were obtained in organic acids compared to mineral acids, and acetic acid was selected as the supporting electrolyte for further studies. Conditions were optimized, and it was possible to determine trace concentrations of Pb by differential pulse anodic stripping voltammetry. A supporting electrolyte of 4% v/v acetic acid, with a deposition potential of -1.5 V (vs. SCE) and a deposition time of 1100 s, was found to be optimum. A linear range of 2.8 µg/L to 110 µg/L was obtained, with an associated detection limit (3σ) of 2.8 µg/L. A mean recovery of 95.6% (CV=3.9%) was obtained for a tap water sample fortified with 21.3 µg/L.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。