Report of Accidental Anchoring of an Impella Device to a Perceval Bioprosthesis in a Patient

报告一例患者体内 Impella 装置意外锚定于 Perceval 生物假体的情况

阅读:1

Abstract

We herein report a case in which we encountered complications when placing an Impella CP ventricular assist device (catheter-based ventricular assist device) in a patient with a Perceval bioprosthetic valve (sutureless valve). Specifically, the catheter-based ventricular assist device became anchored to the sutureless valve and needed to be removed under cardiopulmonary bypass. (Level of Difficulty: Advanced.).