Enzyme-linked immunosorbent assay for measuring antibodies to pneumococcal polysaccharides for the PNEUMOVAX 23 vaccine: assay operating characteristics and correlation to the WHO international assay

用于检测肺炎球菌多糖抗体的PNEUMOVAX 23疫苗酶联免疫吸附试验:试验操作特性及与WHO国际试验的相关性

阅读:1

Abstract

The Merck pneumococcal (Pn) enzyme-linked immunosorbent assays (ELISAs) for measuring antibodies to 12 serotypes (serotypes 1, 3, 4, 6B, 7F, 8, 9V, 12F, 14, 18C, 19F, and 23F) were validated in 1999. Merck Laboratories developed the Pn assays using 10 microg/ml C polysaccharide, 100 microg/ml Pn polysaccharide (PnPs) 25, and 100 microg/ml PnPs 72 for preadsorption of samples, standards, and controls in order to improve the specificity to the Pn serotypes in the vaccine. The Pn assays utilize postimmunization sera obtained from subjects immunized with PNEUMOVAX 23 as standards for measuring immunoglobulin G concentrations in sera obtained from vaccine clinical trials with adults and infants. This material was calibrated to the Pn reference standard serum, 89SF, subjected to the Merck Pn ELISA adsorbants. Comparisons were made between the Merck Pn assay and the international Pn assay, showing moderate agreement between the two assay formats. This work describes the test procedures and operating characteristics of the Merck Pn assays and the results of experiments performed to compare the Merck Pn ELISAs to the international Pn ELISAs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。