Diagnostic Accuracy of Cystic Lesions Using a Pre-Programmed Low-Dose and Standard-Dose Dental Cone-Beam Computed Tomography Protocol: An Ex Vivo Comparison Study

使用预设低剂量和标准剂量牙科锥形束计算机断层扫描方案诊断囊性病变的准确性:一项离体比较研究

阅读:2

Abstract

BACKGROUND: This study aimed to analyze the diagnostic reliability of radiographic assessment of cystic lesions using a pre-set, manufacturer-specific, low-dose mode compared to a standard-dose dental cone-beam computed tomography (CBCT) imaging protocol. METHODS: Forty pig mandible models were prepared with cystic lesions and underwent both CBCT protocols on an Orthophos SL Unit (Dentsply-Sirona, Bensheim, Germany). Qualitative and quantitative analysis of CBCT data was performed by twelve investigators independently in SIDEXIS 4 (Dentsply-Sirona) using a trial-specific digital examination software tool. Thereby, the effect of the two dose types on overall detectability rate, the visibility on a scale of 1 (very low) to 10 (very high) and the difference between measured radiographic and actual lesion size was assessed. RESULTS: Low-dose CBCT imaging showed no significant differences considering detectability (78.8% vs. 81.6%) and visibility (9.16 vs. 9.19) of cystic lesions compared to the standard protocol. Both imaging protocols performed very similarly in lesion size assessment, with an apparent underestimation of the actual size. CONCLUSION: Low-dose protocols providing confidential diagnostic evaluation with an improved benefit-risk ratio according to the ALADA principle could become a promising alternative as a primary diagnostic tool as well as for radiological follow-up in the treatment of cystic lesions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。