Whole-Transcriptome Profiling on Small FFPE Samples: Which Sequencing Kit Should Be Used?

对小型 FFPE 样本进行全转录组分析:应该使用哪种测序试剂盒?

阅读:2

Abstract

RNA sequencing (RNA-Seq) appears as a great tool with huge clinical potential, particularly in oncology. However, sufficient sample size is often a limiting factor and the vast majority of samples from patients with cancer are formalin-fixed paraffin-embedded (FFPE). To date, several sequencing kits are proposed for FFPE samples yet no comparison on low quantities were performed. To select the most reliable, cost-effective, and relevant RNA-Seq approach, we applied five FFPE-compatible kits (based on 3′ capture, exome-capture and ribodepletion approaches) using 8 ng to 400 ng of FFPE-derived RNA and compared them to Nanostring on FFPE samples and to a reference PolyA (Truseq) approach on flash-frozen samples of the same tumors. We compared gene expression correlations and reproducibility. The Smarter Pico V3 ribodepletion approach appeared systematically the most comparable to Nanostring and Truseq (p < 0.001) and was a highly reproducible technique. In comparison with exome-capture and 3′ kits, the Smarter appeared more comparable to Truseq (p < 0.001). Overall, our results suggest that the Smarter is the most robust RNA-Seq technique to study small FFPE samples and 3′ Lexogen presents an interesting quality−price ratio for samples with less limiting quantities.