Analysis of dose measurement other than the radiation protection during the radiographic examination

对放射检查过程中除辐射防护以外的剂量测量进行分析

阅读:1

Abstract

OBJECTIVES: The study measured the dose on body regions that were not shielded to protect from radiation exposure during the general procedure, with the goal of providing basic radiation dose data for radiological technologists who perform the radiographic examination. MATERIALS AND METHODS: Shooting parts with the phantom were similar to human tissues using general shooting equipment in the general examination room. The scattered rays were measured with the ion chamber. The hand received the highest average radiation dose and the kidney the lowest. The same pattern was evident for the average equivalent dose. The available daily shooting was highest in the anterior/posterior skull, followed by the posterior/anterior chest, abdomen, anterior/posterior spine and extremities. RESULTS: The daily available numbers for the eye were lower than other body regions (6-times, 4-times, 26-times, 3-times and 121-times) and the numbers on the foot were higher than for other regions (73-times, 48-times, 263-times, 39-times and 702-times). CONCLUSIONS: Radiation should be thoroughly blocked by the apron to protect the radiological technologist from the radiation exposure, the proper distance from the irradiation source should be maintained exposure is inevitable and the exposure dose and working environment shall be regularly assessed to ensure minimal exposure dose of the radiological technologist in accordance with the International Commission on Radiological Protection recommendation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。