Sub-milliSievert (sub-mSv) CT colonography: a prospective comparison of image quality and polyp conspicuity at reduced-dose versus standard-dose imaging

亚毫西弗 (sub-mSv) CT 结肠造影:低剂量与标准剂量成像的图像质量和息肉清晰度前瞻性比较

阅读:1

Abstract

OBJECTIVE: To prospectively compare reduced-dose (RD) CT colonography (CTC) with standard-dose (SD) imaging using several reconstruction algorithms. METHODS: Following SD supine CTC, 40 patients (mean age, 57.3 years; 17 M/23 F; mean BMI, 27.2) underwent an additional RD supine examination (targeted dose reduction, 70-90%). DLP, CTDI(vol), effective dose, and SSDE were compared. Several reconstruction algorithms were applied to RD series. SD-FBP served as reference standard. Objective image noise, subjective image quality and polyp conspicuity were assessed. RESULTS: Mean CTDI(vol) and effective dose for RD series was 0.89 mGy (median 0.65) and 0.6 mSv (median 0.44), compared with 3.8 mGy (median 3.1) and 2.8 mSv (median 2.3) for SD series, respectively. Mean dose reduction was 78%. Mean image noise was significantly reduced on RD-PICCS (24.3 ± 19HU) and RD-MBIR (19 ± 18HU) compared with RD-FBP (90 ± 33), RD-ASIR (72 ± 27) and SD-FBP (47 ± 14 HU). 2D image quality score was higher with RD-PICCS, RD-MBIR, and SD-FBP (2.7 ± 0.4/2.8 ± 0.4/2.9 ± 0.6) compared with RD-FBP (1.5 ± 0.4) and RD-ASIR (1.8 ± 0.44). A similar trend was seen with 3D image quality scores. Polyp conspicuity scores were similar between SD-FBP/RD-PICCS/RD-MBIR (3.5 ± 0.6/3.2 ± 0.8/3.3 ± 0.6). CONCLUSION: Sub-milliSievert CTC performed with iterative reconstruction techniques demonstrate decreased image quality compared to SD, but improved image quality compared to RD images reconstructed with FBP. KEY POINTS: • CT colonography dose can be substantially lowered using advanced iterative reconstruction techniques. • Iterative reconstruction techniques (MBIR/PICCS) reduce image noise and improve image quality. • The PICCS/MBIR-reconstructed, reduced-dose series shows decreased 2D/3D image quality compared to the standard-dose series. • Polyp conspicuity was similar on standard-dose images compared to reduced-dose images reconstructed with MBIR/PICCS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。