Methodology to analyse small silicon samples by glow discharge mass spectrometry using a thin wafer mask

利用薄晶圆掩模通过辉光放电质谱法分析微小硅样品的方法

阅读:1

Abstract

Glow discharge mass spectrometry (GDMS) is widely used for trace element analysis of bulk solid samples. The geometry of the GD source limits the minimum size of the sample, which for the instrument used in this work (ThermoElementGD) is 20 mm in diameter. From time to time, there is the need to analyse smaller samples with this technique, and we present here a methodology to analyse samples of 9-20 mm diameter through the use of thin masks. Thin masks have been previously used mostly as secondary cathode for the analysis of non-conducting materials, with hole size smaller than the area of the glow discharge. The use of masks in this work includes the following customization:•The choice of highly-pure Si as mask material, to decrease the chance of interferences with the Si samples.•The use of a hole in the mask of the same size as the discharge area. This implies that the mask material is not sputtered, thus decreasing chances for contamination from the mask itself.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。