Video-rate near infrared tomography to image pulsatile absorption properties in thick tissue

利用视频速率近红外断层扫描技术对厚组织中的脉冲吸收特性进行成像

阅读:1

Abstract

A high frame-rate near-infrared (NIR) tomography system was created to allow transmission imaging of thick tissues with spectral encoding for parallel source implementation. The design was created to maximize tissue penetration through up to 10 cm of tissue, allowing eventual use in human imaging. Eight temperature-controlled laser diodes (LD) are used in parallel with 1.5 nm shifts in their lasing wavelengths. Simultaneous detection is achieved with eight high-resolution, CCD-based spectrometers that were synchronized to detect the intensities and decode their source locations from the spectrum. Static and dynamic imaging is demonstrated through a 64 mm tissue-equivalent phantom, with acquisition rates up to 20 frames per second. Imaging of pulsatile absorption changes through a 72 mm phantom was demonstrated with a 0.5 Hz varying object having only 1% effect upon the transmitted signal. This subtle signal change was used to show that while reconstructing the signal changes in a tissue may not be possible, image-guided recovery of the pulsatile change in broad regions of tissue was possible. The ability to image thick tissue and the capacity to image periodic changes in absorption makes this design well suited for tracking thick tissue hemodynamics in vivo during MR or CT imaging.