Time Dependence of Intra-fractional Motion in Spinal Stereotactic Body Radiotherapy

脊柱立体定向放射治疗中分次内运动的时间依赖性

阅读:1

Abstract

BACKGROUND/AIM: Positional uncertainty in spinal stereotactic body radiotherapy (SBRT) may cause fatal error, therefore, we investigated the intra-fractional spinal motion during SBRT and its time dependency. PATIENTS AND METHODS: Thirty-one patients who received SBRT using CyberKnife were enrolled in the study. 2D kV X-ray spine images in two directions were taken before and during treatment. Image acquisition intervals during treatment were set at 35-60 sec. Automatic image matchings were performed between the reference digital reconstructed radiography (DRR) and live images, and the spinal position displacements were logged in six translational and rotational directions. If the displacements exceeded 2 mm or 1 degree, the treatment beam delivery was interrupted and the patient position was corrected by moving couch, and the couch adjustments were also logged. Based on the information, the time-dependent accumulated translational and rotational displacements without any couch adjustments were calculated. RESULTS: Spinal position displacements in all translational and rotational directions were correlated with elapsed treatment time. Especially, Right-Left displacements of >1 mm and >2 mm were observed at 4-6 and 8-10 min after treatment initiation, respectively. Rotational displacements in the Yaw direction >1° were observed at 10-15 min after treatment initiation. CONCLUSION: The translational and rotational displacements systematically increased with elapsed treatment time. It is suggested that the spine position should be checked at least every 4-6 min or the treatment time should be limited within 4-6 minutes to ensure the irradiation accuracy within the millimeter or submillimeter range.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。