Optimization of a Digestion Method to Determine Total Mercury in Fish Tissue by Cold Vapor Atomic Fluorescence Spectrophotometry

利用冷原子荧光光谱法优化消解方法测定鱼组织中总汞含量

阅读:2

Abstract

Several microwave-assisted digestion methods were tested at the Centro de Estudios Aplicados en Química laboratory in Quito, Ecuador, to determine the accuracy and performance efficiency of the mineralization process for the determination of total mercury in fish tissue by cold vapor atomic fluorescence spectrophotometry. The use of MARSEasyPrep high-pressure vessels, low amounts of reagents (1 cm(3) HNO(3), 1 cm(3) H(2)O(2), and 1 cm(3) HClO(4)), an irradiation temperature of 210 °C, and 35 min of mineralization time resulted in accurate performance, with recoveries of certified reference material DORM-4 between 90.1% and 105.8%. This is better than the Association of Official Analytical Chemists 2015.01 method, which has a reported accuracy of 81%. The repeatability precision and intermediate precision were established at three concentration levels (0.167, 0.500, and 0.833 mg·kg(-1)) and expressed as the percentage of the relative standard deviation ranging from 1.5% to 3.0% and 1.7% to 4.2%, respectively. Further, the method was satisfactorily applied to analyze fortified samples of tilapia (Oreochromis niloticus), with recoveries ranging from 98.3% to 104.3%. The instrumental limits of detection and quantification were 0.118 µg·dm(-3) and 0.394 µg·dm(-3), respectively.