Ocular biometry in dense cataracts: Comparison of partial-coherence interferometry, swept-source optical coherence tomography and immersion ultrasound

致密性白内障眼部生物测量:部分相干干涉法、扫频源光学相干断层扫描和浸入式超声的比较

阅读:3

Abstract

PURPOSE: To assess the axial length (AL) measurement failure rate using partial-coherence interferometry (PCI) and swept-source optical coherence tomography (SS-OCT) in dense cataracts. As a secondary outcome, the SS-OCT biometry was compared to immersion ultrasound. METHODS: This is a prospective cross-sectional and comparative study. Seventy eyes from 70 patients with dense cataracts were enrolled in this study. Dense cataract was defined according to the Lens Opacities Classification System III (LOCS III) scores equal to or more than NO4, NC4, C4, and P3. The failure rate of AL measurement was evaluated using PCI and SS-OCT. Anterior chamber depth (ACD), lens thickness (LT), and AL measurements obtained by SS-OCT were compared with IUS. RESULTS: AL measurement failure rate with PCI was 68.57% and 21.43% with SS-OCT (P = 0.007). AL measurement was achieved in 69.23% of NO4, 66.6% of P3, and 15.3% of mixed cataracts using PCI, while SS-OCT was achieved in 100% of NO4, NO5, P3, and P5 and 76.9% of mixed cataracts. Cortical cataracts alone did not influence AL measurement. Biometric data of ACD, LT, and AL were statistically different comparing US and SS-OCT with a good correlation of AL. CONCLUSION: SS-OCT significantly improves the rate of successful AL measurements when compared to PCI in dense cataracts. The LOCS III clinical cut-off for the use of SS-OCT ocular biometry may well be up to P4 and NO5.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。