A novel simplified method for managing inadvertent tube cut during aurolab aqueous drainage implant surgery for refractory glaucoma

一种用于处理难治性青光眼 Aurolab 房水引流植入术中意外导管断裂的新型简化方法

阅读:2

Abstract

We report a novel simplified method for managing inadvertent tube cut in a patient undergoing the Aurolab aqueous drainage implant (AADI) surgery for refractory neovascular glaucoma. Tube cut occurred while applying the polyglactin ligature suture used to avoid early postoperative hypotony. The short end of the cut tube was removed and the long end reinserted into the base plate of AADI. The surgery was then completed as usual. Following the surgery, the intraocular pressure reduced to 20 mmHg which stabilized and was maintained at 10 mmHg till the last follow-up at 12 months without any glaucoma medication.