Plasmonic ELISA for Sensitive Detection of Disease Biomarkers with a Smart Phone-Based Reader

使用基于智能手机的读取器进行等离子体 ELISA 灵敏检测疾病生物标志物

阅读:8
作者:Quanli Yang, Ruitian Cai, Wei Xiao, Zengfeng Wu, Xia Liu, Yan Xu, Miaomiao Xu, Hui Zhong, Guodong Sun, Qihui Liu, Qiangqiang Fu, Junjian Xiang

Abstract

Serum myoglobin is one of the earliest markers for the diagnosis of acute myocardial infarction. It is, therefore, critical to develop a point-of-care testing technology for myoglobin detection. In this work, we reported a sensitive plasmonic immunoassay-based on enzyme-mediated localized surface plasmon resonance change of gold nanorods for the point-of-care testing detection of myoglobin. In addition, we developed a novel plasmonic immunoassay reader using the ambient light sensor of smart phone to increase the accessibility and utility of the plasmonic immunoassay. The linear detection range of gold nanorods-based plasmonic immunoassay for myoglobin detection was 0.1-1000 ng mL-1 and the limit of detection was 0.057 ng mL-1. Myoglobin in serum samples was also analyzed by the plasmonic immunoassay. The results were significantly correlated with those of conventional enzyme-linked immunosorbent assay. The plasmonic immunoassay, coupled with smart phone-based reader, could be widely used for point-of-care testing application of acute myocardial infarction, especially in the regions with limited technological resources.