In Vitro Bacterial Leakage at the Implant-Abutment Connection of Two Dental Implant Systems with Internal Connection

体外细菌渗漏研究:两种内连接式牙科种植系统在种植体-基台连接处的渗漏情况

阅读:1

Abstract

OBJECTIVES: Bacterial leakage at the implant-abutment interface is one of the main causes of peri-implant inflammation. One of the factors that influences bacterial leakage is the structural design of the interface. Considering the limited studies that have examined slip-joint connections, a comparative study of bacterial leakage was performed on two different systems namely Zimmer (Tapered Screw-Vent, Zimmer Dental) with slip-joint connection and Argon (Konus K3pro, Argon Implants) with conical connection. MATERIALS AND METHODS: Twenty-two implants were selected in 2 groups (11 Zimmer with slip-joint connection, and 11 Argon with conical connection) with similar platforms. Escherichia coli (E. coli) suspension (2 μL) was pipetted into the internal lumen of implants. The abutments were screwed onto the implants with a closing torque of 30 Ncm. The assemblies were placed in culture broth for 6, 24, 48 and 72 h, and 7 and 14 days. The colonies were counted and analyzed by the Mann-Whitney test (a=0.05). RESULTS: Microleakage was observed in 20% of the samples of conical connection group after 6 h to 2 days, and in 50% of the samples in slip-joint connection group after 3 to 7 days. There was a significant difference in bacterial leakage rate between the two implant groups (P<0.001) but no significant difference was seen in bacterial leakage over time (P>0.05). CONCLUSION: Type of connection had a significant effect on bacterial leakage, but the rate of bacterial leakage did not significantly change over time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。