Commissioning of a PTW 34070 large-area plane-parallel ionization chamber for small field megavoltage photon dosimetry

PTW 34070 大面积平行平面电离室在小场兆伏光子剂量测定中的调试

阅读:1

Abstract

PURPOSE: This study investigates a large-area plane-parallel ionization chamber (LAC) for measurements of dose-area product in water (DAP(w) ) in megavoltage (MV) photon fields. METHODS: Uniformity of electrode separation of the LAC (PTW34070 Bragg Peak Chamber, sensitive volume diameter: 8.16 cm) was measured using high-resolution microCT. Signal dependence on angle α of beam incidence for square 6 MV fields of side length s = 20 cm and 1 cm was measured in air. Polarity and recombination effects were characterized in 6, 10, and 18 MV photons fields. To assess the lateral setup tolerance, scanned LAC profiles of a 1 × 1 cm(2) field were acquired. A 6 MV calibration coefficient, N(D)(,w,)(LAC) , was determined in a field collimated by a 5 cm diameter stereotactic cone with known DAP(w) . Additional calibrations in 10 × 10 cm(2) fields at 6, 10, and 18 MV were performed. RESULTS: Electrode separation is uniform and agrees with specifications. Volume-averaging leads to a signal increase proportional to ~1/cos(α) in small fields. Correction factors for polarity and recombination range between 0.9986 to 0.9996 and 1.0007 to 1.0024, respectively. Off-axis displacement by up to 0.5 cm did not change the measured signal in a 1 × 1 cm(2) field. N(D)(,w,)(LAC) was 163.7 mGy cm(-2) nC(-1) and differs by +3.0% from the coefficient derived in the 10 × 10 cm(2) 6 MV field. Response in 10 and 18 MV fields increased by 1.0% and 2.7% compared to 6 MV. CONCLUSIONS: The LAC requires only small correction factors for DAP(w) measurements and shows little energy dependence. Lateral setup errors of 0.5 cm are tolerated in 1 × 1 cm(2) fields, but beam incidence must be kept as close to normal as possible. Calibration in 10 × 10 fields is not recommended because of the LAC's over-response. The accuracy of relative point-dose measurements in the field's periphery is an important limiting factor for the accuracy of DAP(w) measurements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。