New, Optimized Skin Calorimeter Version for Measuring Thermal Responses of Localized Skin Areas during Physical Activity

全新优化版皮肤热量计,用于测量运动过程中局部皮肤区域的热反应

阅读:4

Abstract

We present an optimized version of the skin calorimeter for measuring localized skin thermal responses during physical activity. Enhancements include a new holding system, more sensitive thermopiles, and an upgraded spiked heat sink for improved efficiency. In addition, we used a new, improved calorimetric model that takes into account all the variables that influence the measurement process. Resolution in power measurement is 1 mW. Performance tests under air currents and movement disturbances showed that the device maintains high accuracy; the deviation produced by these significant disturbances is less than 5%. Human subject tests, both at rest and during exercise, confirmed its ability to accurately measure localized skin heat flux, heat capacity, and thermal resistance (less than 5% uncertainty). These findings highlight the calorimeter's potential for applications in sports medicine and physiological studies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。