Validation of an Ion-Pair Reverse Phase High-Performance Liquid Chromatography Method for the Detection of Major Components and Related Substances in Diquafosol Sodium Eye Drops

离子对反相高效液相色谱法检测地夸磷索钠滴眼液中主要成分及有关物质的验证

阅读:6
作者:Jingmin Wu, Huina Zhang, Haiming Zhao, Benkai Qin, Tingting Lou, Yonghua Yu, Lihong Huang, Jiangmin Cheng, Hangyu Zhao

Abstract

A simple, feasible, isocratic elution, and stable reversed-phase high performance liquid chromatography method was established and verified. The chromatographic conditions are as follows: EF-C18H, 4.6 × 250 mm, 5 μm column; column temperature 30 °C; for the mobile phase 27.2 g of KH2PO4 and 8.5 g of tetrabutylammonium hydrogen sulfate were taken, 2500 mL of water was added to dissolve, and the pH was adjusted to 6.7 with phosphoric acid:methanol solution with a ratio of 84:16 (V:V). The flow rate was 1.0 mL/min; the injection volume was 10 μL; and the wavelength was 262 nm. According to the current ICH guidelines, the developed method was verified, and the system suitability, specificity, LOD, LOQ, linearity, range, accuracy, repeatability, durability, and solution stability of the proposed method were verified. The validation results demonstrated that the LOQ for the method was 0.05% and the LOD was 0.02%. The content was detected within the concentration range of 300 to 900 μg/mL. The relationship between concentration and measurement was linear, with an r2 of >0.999. The concentration of impurities ranged from 0.3 to 4.5 μg/mL. A good linear correlation was observed within the range of g/mL, with a coefficient of determination r2 greater than 0.999. The accuracy and repeatability met the specified criteria.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。