The use of a personal glucose meter for detecting procalcitonin through glucose encapsulated within liposomes

使用个人血糖仪通过脂质体包裹的葡萄糖检测降钙素原

阅读:6
作者:Fida'a Alshawawreh, Fabio Lisi, Nicholas Ariotti, Padmavathy Bakthavathsalam, Tania Benedetti, Richard D Tilley, J Justin Gooding

Abstract

Herein, a glucose meter-based immunosensing platform is developed that allows the quantification of procalcitonin (PCT) in whole blood samples. PCT is a biomarker for sepsis and its early detection would improve the safety of the patient, as the diagnostic process will be easier and faster. The method employs liposomes with encapsulated glucose as a signal generation tag, which are then used in a sandwich immunoassay by conjugating an antibody to the liposome. The optimal liposomes' size and concentration of encapsulated glucose is determined experimentally to be 200 nm and 27.8 mM, respectively. Upon the addition of a surfactant (Triton X-100), the glucose is released and a signal is detected with a personal glucose meter (PGM). This signal is directly proportional to the concentration of the PCT in the sample. The dynamic range of the assay developed was 0.153-15.38 nM, and could allow the detection of PCT as low as 0.15 nM. The assay showed a high selectivity toward PCT against other proteins such as C-reactive protein and human serum albumin and good reproducibility. This assay was able to quantitatively determine the amount of PCT in whole blood samples at clinically-relevant concentrations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。