Requirements for discrete actuator and segmented wavefront correctors for aberration compensation in two large populations of human eyes

针对两大类人眼像差补偿,对离散致动器和分段式波前校正器的要求

阅读:1

Abstract

Numerous types of wavefront correctors have been employed in adaptive optics (AO) systems for correcting the ocular wavefront aberration. While all have improved image quality, none have yielded diffraction-limited imaging for large pupils (>/=6 mm), where the aberrations are most severe and the benefit of AO the greatest. To this end, we modeled the performance of discrete actuator, segmented piston-only, and segmented piston/tip/tilt wavefront correctors in conjunction with wavefront aberrations measured on normal human eyes in two large populations. The wavefront error was found to be as large as 53 microm, depending heavily on the pupil diameter (2-7.5 mm) and the particular refractive state. The required actuator number for diffraction-limited imaging was determined for three pupil sizes (4.5, 6, and 7.5 mm), three second-order aberration states, and four imaging wavelengths (0.4, 0.6, 0.8, and 1.0 microm). The number across the pupil varied from only a few actuators in the discrete case to greater than 100 for the piston-only corrector. The results presented will help guide the development of wavefront correctors for the next generation of ophthalmic instrumentation.