Clinical comparisons of two free light chain assays to immunofixation electrophoresis for detecting monoclonal gammopathy

两种游离轻链检测方法与免疫固定电泳法在检测单克隆丙种球蛋白病方面的临床比较

阅读:1

Abstract

Free light chains (FLCs) are useful biomarkers for the diagnosis and monitoring of various plasma cell dyscrasias. One hundred fifty-seven samples from 120 patients for screening or monitoring of monoclonal gammopathy (MG) were included. The new N Latex FLC assays (Siemens Healthcare Diagnostics GmbH, Germany) were compared with the Freelite FLC assays (The Binding Site Ltd., UK) and the results were analyzed with those of immunofixation electrophoresis (IFE). The Freelite FLC assay showed significantly wider assay ranges than the N Latex FLC assay. The correlation coefficients of the two FLC kappa (κ) assays, lambda (λ) assays, and the κ/λ ratio were 0.9792, 0.8264, and 0.9064, respectively. The concordance rate was 84.7% for the FLC κ assays, 79.6% for FLC λ, and 89.2% for the κ/λ ratio. The clinical sensitivity and specificity of the κ/λ ratios were 72.2% and 93.6% for the Freelite assay and 64.6% and 100% for the N Latex FLC assay. Two FLC assays showed good correlations and concordance. However, the clinical sensitivity of the κ/λ ratio was higher in the Freelite FLC assays; clinical specificity was higher in the N Latex FLC assay. Both FLC assays seem to have limited clinical utility in detecting MG in certain clinical settings.