Identifying and minimizing primary sources of temporal broadening in online affinity micro free-flow electrophoresis

识别并最大限度地减少在线亲和微自由流电泳中时间展宽的主要来源

阅读:1

Abstract

Affinity micro free-flow electrophoresis (μFFE) takes advantage of continuous separations to enable real-time measurements in biological systems. Competitive immunoassay reagents are mixed online and streamed into the μFFE device, where bound and free labelled species are separated in real-time based on their electrophoretic mobility. In the current manuscript we optimized the affinity μFFE platform to minimize the response time of the assay. Analyte response time was improved by (1) mitigating temporal broadening as analyte transited the fluidic system and (2) reducing surface adsorption. Fast iteration of the microfluidic parameters enabled optimization of temporal broadening due to longitudinal diffusion and parabolic flow. Minimizing incubation capillary length while ensuring adequate equilibration time for the competitive immunoassay improved rise times from 25-seconds to 9-seconds, a 2.7-fold improvement in temporal response. The response time of the assay worsened when immunoassay reagents were incorporated due to peptide surface adsorption. Dynamic coating with 0.005% w/v bovine serum albumin was shown to mitigate surface adsorption without interfering with the binding and separation of the immunoassay components, resulting in 2.8-fold improvement in overall signal and 2-fold improvement in temporal response, which improved from 61 seconds to 30 seconds.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。