Rapid detection of Group B Streptococcus (GBS) from artificial urine samples based on IFAST and ATP bioluminescence assay: from development to practical challenges during protocol testing in Kenya

基于 IFAST 和 ATP 生物发光分析法快速检测人工尿液样本中的 B 组链球菌 (GBS):从开发到肯尼亚协议测试期间的实际挑战

阅读:4
作者:Bongkot Ngamsom, Ernest Apondi Wandera, Alexander Iles, Racheal Kimani, Francis Muregi, Jesse Gitaka, Nicole Pamme

Abstract

We report the rapid detection (20 min) of Streptococcus agalactiae, Group B Streptococcus (GBS) employing on-chip magnetic isolation of GBS based on immiscible filtration assisted by surface tension (IFAST), followed by detection of the isolated GBS using an adenosine triphosphate (ATP) bioluminescence assay. Up to 80% GBS cells were isolated from spiked artificial urine samples with linear responses of bioluminescence signals from isolated cells at 2.3 × 102-9.1 × 105 CFU mL-1, demonstrating great promise for point-of-care detection of pathogenic bacteria in screening urine samples from pregnant women. Practical challenges during initial testing of the developed protocol with urine samples in Kenya are also described.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。