Heartbeat Cycle Length Detection by a Ballistocardiographic Sensor in Atrial Fibrillation and Sinus Rhythm

利用弹道心动描记传感器检测房颤和窦性心律中的心跳周期长度

阅读:1

Abstract

BACKGROUND: Heart rate monitoring is especially interesting in patients with atrial fibrillation (AF) and is routinely performed by ECG. A ballistocardiography (BCG) foil is an unobtrusive sensor for mechanical vibrations. We tested the correlation of heartbeat cycle length detection by a novel algorithm for a BCG foil to an ECG in AF and sinus rhythm (SR). METHODS: In 22 patients we obtained BCG and synchronized ECG recordings before and after cardioversion and examined the correlation between heartbeat characteristics. RESULTS: We analyzed a total of 4317 heartbeats during AF and 2445 during SR with a correlation between ECG and BCG during AF of r = 0.70 (95% CI 0.68-0.71, P < 0.0001) and r = 0.75 (95% CI 0.73-0.77, P < 0.0001) during SR. By adding a quality index, artifacts could be reduced and the correlation increased for AF to 0.76 (95% CI 0.74-0.77, P < 0.0001, n = 3468) and for SR to 0.85 (95% CI 0.83-0.86, P < 0.0001, n = 2176). CONCLUSION: Heartbeat cycle length measurement by our novel algorithm for BCG foil is feasible during SR and AF, offering new possibilities of unobtrusive heart rate monitoring. This trial is registered with IRB registration number EK205/11. This trial is registered with clinical trials registration number NCT01779674.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。