In-silico and in-vitro investigation of a photonic monitor for intestinal perfusion and oxygenation

利用计算机模拟和体外实验研究用于肠道灌注和氧合的光子监测器

阅读:1

Abstract

The quantification of visceral organ oxygenation after trauma-related systemic hypovolemia and shock is critical to enable effective resuscitation. In this work, a photoplethysmography-based (PPG) sensor was specifically designed for probing the perfusion and oxygenation condition of intestinal tissue with the ultimate goal to monitor patients post trauma to guide resuscitation. Through Monte Carlo modeling, suitable optofluidic phantoms were determined, the wavelength and separation distance for the sensor was optimized, and sensor performance for the quantification of tissue perfusion and oxygenation was tested on the in-vitro phantom. In particular, the Monte Carlo simulated both a standard block three-layer model and a more realistic model including villi. Measurements were collected on the designed three layer optofluidic phantom and the results taken with the small form factor PPG device showed a marked improvement when using shorter visible wavelengths over the more conventional longer visible wavelengths. Overall, in this work a Monte Carlo model was developed, an optofluidic phantom was built, and a small form factor PPG sensor was developed and characterized using the phantom for perfusion and oxygenation over the visible wavelength range. The results show promise that this small form factor PPG sensor could be used as a future guide to shock-related resuscitation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。