Non-Contact Radiofrequency Inductive Sensor for the Dielectric Characterization of Burn Depth in Organic Tissues

用于表征有机组织烧伤深度的非接触式射频电感传感器

阅读:1

Abstract

A flat circular transmission line-based 300 MHz resonator was implemented for the non-contact assessment of burn depths in biological tissues. Used as a transmit-and-receive sensor, it was placed at a 2 mm distance from organic material test samples (pork fillet samples) which were previously burned on their surface in various heating conditions involving different temperatures, durations, and procedures. Data extracted from the sensor by means of a distant monitoring coil were found to clearly correlate with the depth of burn observed in the tissue samples (up to 40% sensor output changes for a 7 mm burn depth) and with the heating conditions (around 5% sensor output changes observed in samples burned with identical heating procedures but at two different temperatures-75 °C and 150 °C-and around 40% sensor output changes observed between samples heated at the same temperature but with different heating procedures). These results open the way for the development of easy-to-implement assessment and monitoring techniques for burns, e.g., integrated in wearable medical dressing-like monitoring devices.