Array-based high-intensity focused ultrasound therapy system integrated with real-time ultrasound and photoacoustic imaging

基于阵列的高强度聚焦超声治疗系统,集成实时超声和光声成像技术

阅读:1

Abstract

High-intensity focused ultrasound (HIFU) is a promising non-invasive therapeutic technique in clinical applications. Challenges in stimulation or ablation HIFU therapy are to accurately target the treatment spot, flexibly deliver or fast-move focus points in the treatment region, and monitor therapy progress in real-time. In this paper, we develop an array-based HIFU system integrated with real-time ultrasound (US) and photoacoustic (PA) imaging. The array-based HIFU transducer can be dynamically focused in a lateral range of ∼16 mm and an axial range of ∼40 mm via electronically adjusting the excitation phase map. To monitor the HIFU therapy progress in real-time, sequential HIFU transmission, PA imaging, PA thermometry, and US imaging are implemented to display the dual-modal images and record the local temperature changes. Co-registered dual-modal images show structural and functional information and thus can guide the HIFU therapy for precise positioning and dosage control. Besides therapy, the multi-element HIFU transducer can also be used to acquire US images to precisely align the imaging coordinates with the HIFU coordinates. Phantom experiments validate the precise and dynamic steering capability of HIFU ablation. We also show that dual-modal imaging can guide HIFU in the designated region and monitor the temperature in biological tissue in real-time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。