Impact of scatter on double-pass image quality and contrast sensitivity measured with a single instrument

散射对双程成像质量和对比度敏感度的影响(使用单台仪器测量)

阅读:1

Abstract

We compared objective Double-Pass (DP) image quality data with subjective visual parameters measured within the same modified instrument for different amounts of scatter. The original DP imaging channel of a clinical instrument was maintained intact and two additional channels were included, one for visual testing and another for tear film (TF) imaging by using a retro-illumination technique. Contrast sensitivity (CS) was compared with measurements of the Objective Scattering Index (OSI) obtained from DP retinal images corresponding to different scatter levels induced by pre-defined filters. OSI values were correlated with the change in CS for different spatial frequencies measured with the same instrument. Since TF and DP images were recorded at the same rate, this provided additional information about the dynamic spatial stability of the tear film. This new DP instrument has been proven to provide accuracy and repeatability, and to be suitable for clinical diagnosis, with a complete evaluation of the eye's performance by a simultaneous objective and subjective assessment under the same experimental conditions.