Miniaturized, high numerical aperture confocal fluorescence detection enhanced with pyroelectric droplet accumulation for sub-attomole analyte diagnosis

微型化、高数值孔径共聚焦荧光检测结合热释电液滴聚集技术,用于亚阿托摩尔级分析物诊断

阅读:1

Abstract

To meet the growing demand for early fatal disease screening among large populations, current fluorescence detection instruments aiming at point-of-care diagnosis have the tendency to be low cost and high sensitivity, with a high potential for the analysis of low-volume, multiplex analytes with easy operation. In this work, we present the development of a miniaturized, high numerical aperture confocal fluorescence scanner for sub-micro-liter fluid diagnosis. It is enhanced with high-rate analyte accumulation using a pyroelectro-hydrodynamic dispensing system for generating tiny, stable sample droplets. The simplified confocal fluorescence scanner (numerical aperture 0.79, working distance 7.3 mm) uses merely off-the-shelf mass-production optical components. Experimental results show that it can achieve a high-sensitive, cost-efficient detection for sub-micro-liter, low-abundant (0.04 µL, 0.67 attomoles) fluid diagnosis, promising for point-of-care diagnosis.