Monte carlo simulation of an X-ray pixel beam microirradiation system

X射线像素束微照射系统的蒙特卡罗模拟

阅读:1

Abstract

Monte Carlo simulations are used in the development of a nanotechnology-based multi-pixel beam array small animal microirradiation system. The microirradiation system uses carbon nanotube field emission technology to generate arrays of individually controllable X-ray pixel beams that electronically form irregular irradiation fields having intensity and temporal modulation without any mechanical motion. The microirradiation system, once developed, will be incorporated with the micro-CT system already developed that is based on the same nanotechnology to form an integrated image-guided and intensity-modulated microirradiation system for high-temporal-resolution small animal research. Prospective microirradiation designs were evaluated based on dosimetry calculated using EGSnrc-based Monte Carlo simulations. Design aspects studied included X-ray anode design, collimator design, and dosimetric considerations such as beam energy, dose rate, inhomogeneity correction, and the microirradiation treatment planning strategies. The dosimetric properties of beam energies between 80-400 kVp with varying filtration were studied, producing a pixel beam dose rate per current of 0.35-13 Gy per min per mA at the microirradiation isocenter. Using opposing multi-pixel-beam array pairs reduces the dose inhomogeneity between adjacent pixel beams to negligible levels near the isocenter and 20% near the mouse surface.