An evaluation of M2+ interference correction approaches associated with As and Se in ICP-MS using a multi-day dataset along with ICP-MS/MS/HR-ICP-MS based analysis and hierarchical modeling as a means of assessing bias in fortified drinking waters and single component matrices

使用多日数据集以及基于 ICP-MS/MS/HR-ICP-MS 的分析和分层建模来评估与 ICP-MS 中的 As 和 Se 相关的 M2+ 干扰校正方法,作为评估强化饮用水和单组分基质中偏差的手段

阅读:5
作者:Skyler W Smith, Roy W Martin, Nicole Hanks, Patricia A Creed, Kasey Kovalcik, Robert A Wilson, Kevin Kubachka, Judith A Brisbin, Julio A Landero Figueroa, John T Creed

Abstract

Three 1212<math> <mrow><mfrac><mn>1</mn> <mn>2</mn></mfrac> </mrow> </math> mass oriented rare earth element (REE) M2+ correction approaches (fixed factor, a dual internal standard, and an in-sample) are evaluated for use in an ICP-MS environmental method update. The multi-variant-based evaluation includes analyzing the same 19 REE-fortified matrices on eight different days over a two-month period using two instrument tunes. These REE-fortified matrices were also analyzed using HR-ICP-MS and ICP-MS/MS to estimate the reference value for use in the principal component analysis (PCA) and hierarchical modeling evaluation. A fixed factor is unable to compensate for matrix and mass dependent drift and because of this it generates the largest across matrix, tune, and day 95th percent confidence bounds for the REE corrections on both As (1.1 ppb) and Se (23 ppb) using samples fortified with 100 ppb Nd, Sm & Gd. The PCA analysis indicated that M2+ ions cluster together across matrix, tune and day better than M1+ and these tighter correlations are reflected in reduced 95th percentile confidence bounds for dual M2+ internal standards (M2+; As = 0.3 ppb; Se = 5.4 ppb; n = 704) relative to M1+ internal standards (M1+; As = 0.6 ppb; Se = 12.0 ppb; n = 1056). The use of an in-sample M2+ correction produced comparable 95th percent confidence bounds (As = 0.2 ppb; Se = 3.4 ppb; n = 352) relative to the M2+ internal standard approaches. Finally, the hierarchical modeling indicated M2+ ions as internal standards tend to minimize the across day variability induced by cone changes and the daily reoccurring matrix shifts in the M2+/M1+ ratio associated with 250 ppm matrices of Na, Ca, and Mg. This internal standard driven reduction in variability can be beneficial in compliance monitoring methods.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。